How-To Guide

Lab Film Heat Seal Testing Procedure: ASTM F2029 Step-by-Step Protocol

ASTM F2029 is the standardized procedure for making laboratory heat seals that are used as input to ASTM F88 strength testing. Following this protocol precisely is the foundation of reproducible, inter-lab-comparable heat seal data.

Equipment Setup: Heat Seal Tester Calibration and Parameter Configuration

Before testing, verify instrument calibration: temperature accuracy at planned test points (ASTM F2029 requires ±1°C; KHT STH models provide ±0.5°C — verify with calibrated reference thermocouple at least annually); jaw force calibration (verify with calibrated load cell); jaw surface condition (clean, flat, no contamination from previous tests). Allow instrument to equilibrate at set temperature for at least 10 minutes before making test seals. For STH-3A gradient testing: allow all 5 zones to stabilize independently — verify set temperature vs. displayed temperature at each zone.

Sample Preparation: Film Specimen Dimensions, Conditioning, and Orientation

Cut film specimens to 150–200mm in the machine direction (MD) × 15mm or 25mm in the transverse direction (TD). If the film has a specific machine direction (e.g., oriented films), cut specimens consistently with MD orientation documented. Condition specimens at 23°C ±2°C, 50% ±5% RH for minimum 24 hours before sealing (or as specified by the relevant product standard). When sealing: orient with sealant layers facing each other (sealant-to-sealant for self-sealing films) or as required by the application (e.g., film sealant to foil lidstock).

Running the Test: Temperature, Dwell Time, and Pressure Settings

Per ASTM F2029: Set temperature to target value. Allow 10 minutes equilibration. Set dwell time and jaw pressure. Insert specimen under jaw so that the intended seal zone is centered under the jaw. Close jaw — ASTM F2029 defines jaw closure as the moment of full contact under set pressure. Hold for programmed dwell time. Open jaw. Remove specimen. Record: seal temperature (°C), jaw pressure (N or N/cm²), dwell time (seconds), specimen ID, lot/batch of film, operator, date, instrument ID and calibration date. For gradient testing (STH-3A): all 5 positions are sealed simultaneously — record gradient settings (all 5 temperatures).

Post-Seal Conditioning and Peel Strength Measurement Per ASTM F88

After sealing: condition specimens at 23°C ±2°C, 50% ±5% RH for minimum 1 hour (or 24 hours for pharmaceutical applications or as required by specification). Do not stack or compress sealed specimens during conditioning. After conditioning: cut sealed specimens to 15mm or 25mm width as required. Mount in tensile tester, set crosshead speed to 250–300mm/min, select peel angle (Technique A, B, or C per ASTM F88 and application). Record peak force (N) and calculate N/15mm (or N/25mm). Report as mean ± standard deviation for replicate specimens (minimum 3 replicates per ASTM F88).

Data Recording and Inter-Lab Comparability Best Practices

For inter-lab comparable results, document: instrument make/model/serial number; calibration status and date; film lot number and supplier; conditioning environment (temperature/humidity log); sealing parameters (temperature, force, dwell for all positions); peel testing parameters (speed, angle, specimen width); results (individual values + mean + SD); failure mode (adhesive, cohesive, film break — per ASTM F88 definitions). The most common cause of inter-lab discrepancy in heat seal testing is undocumented or uncontrolled sealing parameters — particularly dwell time variation and uncalibrated temperature. ASTM F2029 exists specifically to eliminate this.

Troubleshooting: Cold Seals, Delamination, and Film Degradation

Cold seals (strength below minimum): temperature below SIT; insufficient dwell time; contamination in seal zone; sealant-to-substrate (wrong orientation). Resolution: use STH-3A gradient to confirm SIT; verify dwell time programming; clean jaw surfaces; verify specimen orientation. Delamination (substrate layers separating, not seal opening): temperature too high; dwell time too long; excessive jaw pressure. Resolution: reduce temperature or dwell; recalibrate jaw pressure. Film degradation (burns, discoloration, structural failure): temperature too high for material. Resolution: reduce temperature; verify material's specified sealing range.

Ready to Apply This in Your Lab?

Tell us your packaging material and testing standard — we'll recommend the right KHT STH model.

Get a Quote →

Frequently Asked Questions

Questions About Heat Seal Testing?

Our engineers answer specification, compliance, and selection questions. Factory-direct pricing, 24-hour response.

Contact KHT Instrument